Stabilitas Whole Blood Control Tiga Level Yang Disimpan Pada Suhu 2-8oC Menggunakan Alat Hematology Analyzer

Authors

  • Adela Caesaria Politeknik Kesehatan Kemekes Surabaya
  • Anik Handayati Politeknik Kesehatan Kemekes Surabaya
  • Ayu Puspitasari Politeknik Kesehatan Kemekes Surabaya

DOI:

https://doi.org/10.26630/jak.v12i1.3492

Keywords:

Stability. Complete blood count, Complete blood cells, Hematology

Abstract

Quality assurance is important to ensure that the quality system runs properly and guarantees the results of laboratory tests, one of which is the inspection of control materials. The control material requirement is to have long stability, blood-like characteristics and low price. From these conditions, commercial control materials have a longer stability than fresh blood but are expensive. In practice, commercial control materials are still used even after the expiry date of the materials has passed. This is because the availability of materials is often delayed, the procurement of materials takes a long time and the price is relatively expensive. Therefore, this study aims to determine the stability of commercial whole blood control stored at 2-8oC. This research is an experimental analytic study with a cross sectional research design. The sample in this study was commercial whole blood control with 3 levels, namely normal, low abnormally and abnormally high levels which were stored at 2-8oC and checked every week for 10 weeks. Based on the test results obtained p-value for each parameter at all levels on the parameters of the number of erythrocytes, leukocyte count, hematocrit, hemoglobin and platelet count. Based on the results, it can be concluded that there is no change in the stability of whole blood control with storage for 10 weeks at a temperature of 2-8oC on the parameters of abnormally low and normal leukocyte count levels, abnormally low and abnormally high hematocrit levels, and all levels on the parameters of leukocyte count, hemoglobin and platelet count. Meanwhile, in the normal level of hematocrit parameters and abnormally high leukocyte count, there is a change in stability.

References

Gundersen. (2018). Sysmex XN-450/XN-550 Complete Blood Count and Parameters – Whole Blood. Standard Operating Procedure, 1–25.

Handayani, E., Rahmawati, A., & Niken. (2019). Faktor-Faktor Yang Mempengaruhi Kejadian Preeklampsia Di RSUD Wates Kabupaten Kulonprogo Tahun 2019. Poltekkes Kemenkes Yogyakarta.

Laelasari, T. (2020). Verifikasi Metode Pemeriksaan Hematologi Lengkap Dengan Alat Hematology Analyzer Pada Rumah Sakit Di Kabupaten Bandung [Politeknik Kesehatan Kemenkes Bandung]. http://repo.poltekkesbandung.ac.id/665/

Mardiana, & Rahayu, I. G. (2017). Bahan Ajar Teknologi Laboratorium Medis (TLM) Pengantar Laboratorium Medik. Badan Pengembangan dan Pemberdayaan Sumber Daya Manusia Kesehatan Kemeterian Kesehatan Republik Indonesia. http://bppsdmk.kemkes.go.id/pusdiksdmk/wp-content/uploads/2017/11/Pengantar-Laboratorium-Medik-SC.pdf

Mariah. (2017). Analisa Kontrol Kualitas Internal Pemeriksaan Hemoglobin, Leukosit Dan Trombosit Pada Alat Hematology Analyzer Di Laboratorium “X†Wilayah Samarinda. Sekolah Tinggi Ilmu Kesehatan Wiyata Husada Samarinda.

Nirwani, Hartiti, T., & Faruq, Z. H. (2018). Analisis Akurasi Dan Presisi Alat Hematology Analyzer ABX Pentra XL 80 Di Laboratorium Rumah Sakit Roemani Muhammadiyah Semarang. Analisis Akurasi Dan Presisi Alat Hematology Analyzer Abx Pentra Xl 80 Di Laboratorium Rumah Sakit Roemani Muhammadiyah Semarang.

Oktarini. (2018). Stabilitas Objek Uji Profisiensi Hematologi Terhadap Variasi Suhu Dan Waktu Pada Program Uji Profisiensi Di Laboratorium Kesehatan Provinsi Jawa Barat [Poltekkes Kemenkes Bandung]. https://kink.onesearch.id/Record/IOS6083.3133/Preview#toc

Permenkes, R. (2010). Peraturan Menteri Kesehatan Republik Indonesia Nomor 411/Menkes/Per/III/2010 Tentang Laboratorium Klinik. In Pusat Komunikasi Publik Departemen Kesehatan. https://pelayanan.jakarta.go.id/download/regulasi/peraturan-menteri-kesehatan-nomor-411-tahun-2010-tentang-laboratorium-klinik.pdf

Rahmanitarini, A., Hernaningsih, Y., & Indrasari, Y. N. (2019). The Stability of Sample Storage For Complete Blood Count (CBC) Toward The Blood Cell Morphology. Bali Medical Journal, 8(2), 482. https://doi.org/10.15562/bmj.v8i2.1369

Ranggaeni, L. (2016). Gambaran Hasil Pemeriksaan Bahan Kontrol Buatan Sendiri Untuk Hematology Analyzer. Sekolah Tinggi Ilmu Kesehatanmuhammadiyah Ciamis.

Rohmah, A. (2019). Gambaran Pemantapan Mutu Hematologi (Pemeriksaan Hemoglobin dan Leukosit) di Laboratorium Puskesmas Wilayah Kerja Kabupaten Mojokerto. Journal of Chemical Information and Modeling, 53(9), 1689–1699.

Syamsyuddin. (2017). SOP Hematologi Analyzer SYSMEX XP -100. https://pdfcoffee.com/hematologi-analyzer-sysmex-xp-100-pdf-free.html

Sysmex. (2017). Sysmex Educational Enhancement and Development. SEED Hematologia, March.

Tuna, H., & Widyaningsih, A. (2017). Perbandingan antara bahan kontrol komersial merk diasys-trulab n dengan siemens-biorad level 1 terhadap akurasi untuk pemeriksaan glukosa, kolesterol dan asam urat. Jurnal Wiyata Penelitian Sains Dan Kesehatan, 3(1), 85–91. http://ojs.iik.ac.id/index.php/wiyata/article/view/75/74

Tuntun, M., Wulan, W. S., Setiawan, D., & Nuryati, A. (2018). Bahan Ajar Teknologi Laboratorium Medik (TLM) Kendali Mutu. Badan Pengembangan Dan Permberdayaan Sumber Daya Manusia Kesehatan Kemetetian Kesehatan Republik Indonesia. http://bppsdmk.kemkes.go.id/pusdiksdmk/wp-content/uploads/2018/09/Kendali-Mutu_SC.pdf

Vis, J. Y., & Huisman, A. (2016). Verification and Quality Control of Routine Hematology Analyzers. International Journal of Laboratory Hematology, 38, 100–109. https://doi.org/10.1111/ijlh.12503

Downloads

Published

2023-06-26